Operator's Manual
117
Troubleshooting
,
continued
8. Troubleshooting
Battery
In case you experience one of the following issues:
•
The analyzer is turned off as soon as the power cable is disconnected
from the outlet
•
The analyzer is turned off during a short power outage
•
The battery does not recharge.
1. If the analyzer is on, please turn it off.
2. Check if the connection between the analyzer and the outlet is loose.
3. Stabilize any loose power connection.
4. Turn the power switch on.
5. The analyzer should be turned on and the battery starts charging.
6. If the analyzer is not on, turn off the power switch again.
7. Charge the battery for approximately 10 minutes.
8. Turn on the analyzer’s power switch back on.
9. Check the battery level while the analyzer is on.
10. In case the charging level does not go up, contact your reseller or ©
i-SENS Customer Service center.
11. The analyzer will operate properly as long as it is connected with the
power cable until the battery is replaced.
•
If the “
Battery is low
” appears.
1. Check if the power connection between the analyzer and the power is
loose.
2. Fix any loose power connection.
Содержание i-SmartCare 10
Страница 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Страница 3: ......
Страница 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Страница 5: ......
Страница 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Страница 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Страница 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Страница 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Страница 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Страница 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Страница 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...