Troubleshooting
,
continued
8. Troubleshooting
Power
•
Power connected icon
is not indicated while power cable is
plugged into the analyzer.
1. Check if the connection between the analyzer and the power cable is
loose.
2. Fix the loose power connection.
In case you experience one of the following issues:
•
The analyzer is turned off.
•
The analyzer is not turning on.
1. If the analyzer’s power switch is on, turn it off.
2. Check if the connection between the analyzer and the power outlet is
loose. Fix the loose power connection.
3. Turn on the analyzer.
4. In case the analyzer is not turning on, turn off the power switch again.
5. Charge the battery for approximately 10 minutes.
6. Turn on the power switch of the analyzer again.
Содержание i-SmartCare 10
Страница 1: ...PZA1O7619 REV7 2020 12 Operator s Manual Operator s Manual...
Страница 3: ......
Страница 4: ...Please take time to read this operator s manual before setting up and operating the analyzer...
Страница 5: ......
Страница 8: ...1 Introduction 1 Introduction Safety 8 Symbols and Terminology 10 Product Description 13...
Страница 27: ...2 Installation 26 i SmartCare 10 Blood Gas Analyzer This page is intentionally left blank...
Страница 31: ...30 i SmartCare 10 Blood Gas Analyzer Screen Menu continued 3 Instrument Settings Menu options...
Страница 68: ...4 Sample Analysis 4 Sample Analysis Start Analysis 68 Blood Sample 69 QC Sample 78...
Страница 86: ...5 Database 5 Database Sample Data 86 QC Data 90 Calibration Data 96 Cartridge Data 97...
Страница 100: ...6 Shutdown 6 Shutdown Analyzer Shutdown 100 Power Recovery 102...
Страница 116: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 116 Error Code 122...