28
www.i-sens.com
www.i-sens.com
29
Step 13
This mode allows you to turn the hyperglycemia indicator (possible
high blood sugar) On or OFF and to select the desired level for
the indicator. You will be alerted any time your test result is higher
than the selected level. On pressing the
◀
or
▶
button, the screen
will display 'On' or 'OFF'. Press the
S
button when 'On' appears
to enter the setting. Press the
◀
or
▶
button until the desired
hyperglycemia level between 120–349 mg/dL (6.7–19.4 mmol/L)
appears. Press and hold the
S
button to confirm the hyperglycemia
level and turn the meter off.
Setting the Hyperglycemia (HI) Indicator
Caution:
Ask your healthcare professional to help you decide what your
hyperglycemia level is before setting your level.
Note:
If the test result is higher than the
pre-set hyperglycemia level, the
meter will display 'HI' on the bottom.
'KETONE' symbol blinks 3 times when
the result is between 241 mg/dL
(13.4 mmol/L) and 600 mg/dL (33.3 mmol/L).