CareSens PRO Control Solution with CareSens PRO Test Strips
KetoSens Control Solution with KetoSens Test Strips
Note:
• Use the CareSens PRO Control Solutions and KetoSens
Control Solutions only with their corresponding test strips.
The control solutions are available for purchase separately.
• Check the expiration date printed on the bottle.
• Make sure your meter, test strips, and control solution are at
room temperature before testing. Control solution tests must
be done at room temperature (20–25 ˚C).
• Before using the control solution, shake the bottle, discard
the first few drops and wipe the tip clean.
• Close the control solution bottle tightly and store at a
temperature between 8–30 ˚C.
You may do a control solution test:
• When you want to practice the test procedure using the control
solution instead of blood,
• When using the meter for the first time,
• Whenever you open a new vial or a new box of test strips,
• If the meter or test strips do not function properly,
• If your symptoms are inconsistent with the blood glucose test
results and you feel that the meter or test strips are not working
properly,
• If you drop or damage the meter.