Control Solution Testing
Step 4
Waiting for the Result
The display segments will rotate
clockwise and a test result will
appear after the meter counts
down from 5 to 1 if you test with
CareSens PRO test strip and
CareSens PRO control solution.
In case of KetoSens test strip and
KetoSens control solution, the test result will appear after the
meter counts down from 8 to 1.
The test result with control solution flag is stored in the memory
but not included in the averages.
2.0-3.8 mmol/L
CareSens PRO Control Solution Range
Control L :
XX–XX mg/dL (X.X–X.X mmol/L)
Control H:
XX–XX mg/dL (X.X–X.X mmol/L)
Note:
The meter may switch off if the control solution sample
is not applied within 2 minutes of the symbol appearing on
the screen. If the meter turns off, remove the strip, reinsert, and
start from step 1.
Step 5
Comparing the Result
Compare the result displayed on the
meter to the range printed on the test
strip vial or box. The result should fall
within the range.
Caution:
The range printed on the test strip vial or box is for the
Control Solution only. It has nothing to do with your blood glucose
or your blood
β
-ketone levels.