12
www.i-sens.com
www.i-sens.com
13
Warning!
• The CareSens PRO test strips should be used with fresh
capillary whole blood samples, or with fresh venous, neonatal
and arterial whole blood samples if drawn by healthcare
professionals. Besides whole blood samples, serum or plasma
samples can affect test results.
• Fresh venous and arterial whole blood specimens containing
the anticoagulants EDTA and Heparin are acceptable.
Iodoacetate or fluoride/oxalate should not be used.
CareSens PRO Blood Glucose Test Strip
The CareSens Dual Blood Glucose Monitoring System measures
blood glucose quickly and accurately. It automatically absorbs
the small blood sample applied to the narrow edge of the strip.
Contact bars
Gently push the test strip, with its
contact bars facing up, into the test
strip port of meter
Confirmation window
Check here to see whether sufficient
blood sample has been applied
Edge to apply blood sample
Apply blood sample here for testing