3
INS0228 July 2020
Therefore, an RLU value is not t he sam e as a m icrobial colony form ing
unit ( CFU) . Since t he lum inom et er is det ect ing t ot al ATP, it is unknown
whet her t he RLU result displayed by t he lum inom et er is due t o t he
det ect ion of m icrobial ATP, residual ATP, or a com binat ion of bot h.
Therefore, a com parison cannot be drawn bet ween RLU values from
ATP and st andard plat e counts ( SPC) ; or rat her RLU does not equal
CFU. True m icrobial cont am inat ion can only be det erm ined by a
st andard plat e count m et hod that can t ake up t o 48 hours for result s.
The lum inom et er is capable of providing result s in 15 seconds offering
a rapid, real- t im e approach t o hygiene/ sanit at ion m onit oring. The
lum inom et er is not m eant to replace m icrobial t est ing. I t can be an
effect ive t ool when incorporat ed int o a Hazard Analysis Crit ical Cont rol
Point ( HACCP) , Good Manufact uring Pract ices ( GMP) , or sanit at ion
plan.
1 .4
Te st De v i ce s a n d A cce sso r i e s
Device
Cat No.
Description
UltraSnap
US2020
Surface total ATP test device
SuperSnap
SUS3000
High sensitivity total ATP test suited
for allergen prevention applications
AquaSnap Total
AQ100X
Water total ATP test device
AquaSnap Free
AQ100FX
Water Free ATP test device
Contamination
Control Kit
CCK-4
Cuvette based water testing kit
MicroSnap
Enterobacteriaceae
MS-EB
Enterobacteriaceae test
MicroSnap
Coliform Test
MS-COLIFORM Coliform test
MicroSnap E. coli
Test
MS-ECOLI
E. coli test
MicroSnap Total
MS-TOTAL
Total Viable Count test
ZymoSnap ALP
ZS
Alkaline Phosphatase test
CrossCheck
CX3000
Acid Phosphatase test
Calibration Kit
PCD4000
Positive and negative rod for
validating luminometer calibration
ATP Positive
Control Kit
CK25
25 vials of freeze-dried ATP to verify
efficacy of ATP test devices
See t he Hygiena websit e for a com plet e list of t est devices,
accessories, and ot her product s.