10
INS0228 July 2020
N OTE: SureTrend soft ware is required for sect ions
3.1
to
3.3
. For
furt her det ails, refer t o t he separat e SureTrend User's Manual locat ed
on t he SureTrend soft ware CD t hat is packaged wit h t he lum inom et er .
3 .1
Use r I D
The lum inom et er can store a list of 201 User I Ds ( User: 0 t o 200) .
Each t est result will be t agged wit h t he current ly select ed User I D.
Select t he Use r I D m enu option, t hen use t he
and
but t ons t o
scroll t hrough t he list of User I Ds, t hen press t he
but t on t o select
a User I D, or t he
but t on to cancel t he select ion.
TI P: Use r : 0 is t he default select ion if no User I D list has been
defined.
3 .2
Se t t i n g u p Pr o g r a m Te st Po i n t s a n d Pass/ Fa i l Li m i t s
The lum inom et er has t he abilit y t o store a list of 5001 unique Program
t est point s ( Prog# 0 t o 5000) , each w it h it s own upper ( ) and lower
( ) pass/ fail result t hreshold values.
Program t est point locat ion nam es can only be program m ed int o t he
lum inom et er t hrough t he SureTrend soft ware. Program pass/ fail
t hresholds can eit her be set - up on t he lum inom et er , or set using t he
SureTrend soft ware. I f a locat ion nam e is ent ered and t hreshold set
in SureTrend, and t hen downloaded t o t he lum inom et er, t he threshold
values can’t be changed via t he lum inom et er . This is a securit y feat ure
and prevent s inadvert ent changes to t he t hresholds aft er being set in
SureTrend.
TI P: Pr o g : 0 has a special funct ion, whereby t he t hreshold values
defined for Pr o g : 0 are used as t he init ial values when set t ing- up t he
ot her Pr o g t hresholds. Set t ing Pr o g : 0 wit h your typical t hreshold
values will save t im e when set t ing- up t he ot her Program s. The
lum inom et er com es wit h pass/ fail t hresholds of 10 RLU for pass and
30 RLU for fail. These pass/ fail t hresholds were det erm ined based on
a com prehensive st udy of ATP, food residue, and m icroorganism
sam ples. To receive t he report or m ore inform at ion on set t ing up
pass/ fail t hresholds cont act Hygiena or visit our Quest ion and Answer
sect ion at
www.hygiena.com
.
To set t he Program result t hresholds on t he lum inom et er , first select
t he PROGRAM m enu opt ion by pressing
. Scroll to “ PROGRAM” and
press
. Then use t he
and
but t ons t o scroll t hrough t he list of
Program s, followed by t he
but t on t o select t he Program t o be
changed: