Hydro-View User Guide HD0531 Rev 2.0.0 51
Unscaled 2 (only shown for selected sensors)
For sensors which support this option, this shows the current Measurement Mode used to
calculate the Unscaled 2 reading. Touch to change. For further details of the Measurement
Modes, refer to the Calibration and Configuration Guide HD0679.
Material Temperature Alarm Low/High
The Material High and Low Limits are used to configure the Material Temperature alarm. If
Digital Input/Output 2 is set to Material Temperature Alarm the output will become active if
the material temperature sensor is above the high limit or below the low limit.
Alarm Mode
Configures which Measurement Mode (Mode F, Mode V, Mode E or Legacy) is used to
calculate the alarm values. The Alarm Mode is only available for sensors with multi
Measurement Mode capabilities. Once configured, the sensor will only calculate the alarm
values using the selected Measurement Mode. The Alarm Mode will also configure which
Measurement Mode is used to calculate the AutoTrack values.
6.6
Factory Settings Screen
The Calibration and Configuration Guide HD0679 should be read and understood
before attempting to change these parameters.
6.6.1
Factory Settings
Figure 34: Factory Settings Screen
Water/Air
These show the Air and Water points used to calculate the Unscaled readings. These are
set in the factory and should not be manually entered.
AutoCal
When fitting a new Sensing Arm to a Hydro-Probe Orbiter, or replacing the ceramic on a
Hydro-Mix the Air and Water factory calibrations need to be updated. However, if the
sensor is installed in a mixer, it is not always possible to manually perform the air and
water readings. To help with this issue an alternative function called AutoCal may be used.
This makes an air reading and then estimates the water reading based on a constant air-
water difference.
When using AutoCal with a Hydro-Probe Orbiter certain arm types require the selection of
the type and length before this calibration starts. The arm type and length is set on the
Digital Inputs/Outputs Screen (Page 46).
Содержание HV04
Страница 4: ...4 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 6: ...6 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 12: ...12 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 14: ...14 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 22: ...Mechanical Installation Chapter 2 22 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 28: ...Electrical Installation Chapter 3 28 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 78: ...Material Calibration Chapter 6 78 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 80: ...Default PIN Codes Appendix A 80 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 82: ...USB Memory Stick file format Appendix B 82 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 84: ...Quick Start Rules Appendix C 84 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 86: ...Frequently Asked Questions Appendix D 86 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 92: ...Document Cross Reference Appendix G 92 Hydro View User Guide HD0531 Rev 2 0 0 ...