Hydro-View User Guide HD0531 Rev 2.0.0 47
Sensor OK
This output will be active if:
•
The frequency reading is between the defined air and water
calibration /-3%
•
The amplitude reading is between the defined air and water
calibration /-3%
•
The temperature of the internal electronics is below the safe
operating limit
•
The temperature of the RF resonator is above it’s safe
operating limit
•
The internal supply voltage is in range
Material Temp: This alarm will be active if the material temperature is outside the
configured high/low limits.
Calibration Out or Range The output will be active if the Unscaled reading, for any of
the Measurement Modes, is more than 3 points above or
below the range of Unscaled values used in the calibration.
This can be used to indicate that another calibration point
could/ should be made.
AutoTrack Stable:
AutoTrack Stable indicates if the sensor reading is stable.
The stability is defined as the deviation of a set amount of
data points. Both the deviation value and the amount of data
used, in seconds, are configurable in the sensor. The output
will be active if the AutoTrack Deviation is below the
AutoTrack Deviation threshold
Arm Type
-
The current type of Sensing Arm fitted if the sensor is a Hydro-Probe Orbiter.
Arm ID
-
The ID number of the connected Hydro-Probe Orbiter Sensing Arm
Test I/O
-
Accesses the I/O Test Screen (see I/O Test Screen, section 6.4).
6.4
I/O Test Screen
The I/O Test screen provides a method of testing the I/O, to ensure that the sensor is
correctly connected to the control system. Be aware that if the control system is running,
manually adjusting the I/O may have unexpected consequences.
Analogue Output Control Sliders
Analogue Output Current
Toggle Switches
Digital I/O Status
Figure 29: I/O Test Screen
1
: - Slider for testing Analogue Output 1. Current mA value is displayed in the box.
2
: - Slider for testing Analogue Output 2 (not available on all sensor models). Current mA
value is displayed in the box.
Содержание HV04
Страница 4: ...4 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 6: ...6 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 12: ...12 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 14: ...14 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 22: ...Mechanical Installation Chapter 2 22 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 28: ...Electrical Installation Chapter 3 28 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 78: ...Material Calibration Chapter 6 78 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 80: ...Default PIN Codes Appendix A 80 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 82: ...USB Memory Stick file format Appendix B 82 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 84: ...Quick Start Rules Appendix C 84 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 86: ...Frequently Asked Questions Appendix D 86 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 92: ...Document Cross Reference Appendix G 92 Hydro View User Guide HD0531 Rev 2 0 0 ...