7 | Operation
HYDAC FILTER SYSTEMS GMBH
54 / 92
BeWa FCU1210 3545830b en-us lq
Fig. 22:
Operating Mode - Measurement ≤ 45 bar
Make sure that the following sequence is observed:
ü
You have performed all of the steps listed in chapter
Putting the FCU into operation [
.
1.
Check the pressure at the measurement location. The
existing pressure must be in the range of 1 … 45 bar /
14 … 650 psi.
WARNING! If the pressure is greater than 45 bar /
650 psi, then use the high-pressure adapter.
2.
Connect the entry hose (color: black) with the INLET
connection of the FCU (1).
Rotate the measurement coupling clockwise (2) to the
connection and tighten it by hand.