Operation | 7
BeWa FCU1210 3545830b en-us lq
59 / 92
7.3 Performing Measurements
To carry out a measurement, observe the following sequence:
ü
Preparing the FCU, see
Putting the FCU into operation
ü
Make the desired settings for the measurement, see Op-
erating/Setting measuring menu
ü
Select a corresponding operating mode for the pressure,
see
Select the operating modes / measurement method
according to the pressure involved [
1.
Switch on the pump at the switch.
The pump conveys the fluid for analysis through the
FCU.
2.
After the set measurement time, the result is shown on
the display, and the status LED lights up green perma-
nently.
3.
Switch the pump off after 4 minutes.
ð
The measurement is complete.