26
14. The different frequency should be adopted for measuring if
the result cannot be confirmed measured by working frequency
with too much interferences. Otherwise, working frequency should
be used for the measurement.
15. If the measurement result cannot be shown long time (1
min) after sending out the measurement instruction, it may be
caused by large capacitance of the specimen or crashing of the
instrument. You need to restart the instrument, reduce the
measurement voltage and make measurement.
16. The instrument has automatic protection feature if it has
short circuit.
17. If the paper is extended out from the paper exit on the top
of the printer, you need to press the button to stop it. The printing
paper could be pulled out.
VIII. Full set of instrument
1. Host
1
2. High-voltage HVx test cable (Red)
1
3. Cx-specific test cable (black)
1
4. High-voltage output test cable (for CVT, dual-core)
1
5. Low-voltage output test cable (for CVT, dual-core)
1
6. 220 V power line
1
7.Ground wire
1
8. Fuse
4