![Huvitz HRK-9000A Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/huvitz/hrk-9000a/hrk-9000a_service-manual_2190272052.webp)
---------------------------------------------------------------------------------------------- Auto Ref/Keratometer HRK-9000A 51
3) If you press the joystick, Dead Pixel beginning of the search
and discovery of the Dead Pixel count is displayed on the screen.
Ten thousand days When Buffer Full output with the message,
DPC compensation will exceed the number of possible Dead
Pixel.
D. C in the same way and then Setup and then to the Dead Pixel.
1) KER CCD's White Pixel
2) KER CCD
’s Black Pixel in
3) COLOR CCD's White Pixel
4) COLOR CCD
’s Black Pixel in
E. After DPC Setup, engineer mode by pressing the SAVE SYS
DATA is stored.
F. HRK-8000A restarts, and were being corrected by the camera
from the screen Dead Pixel Check.
4-1-17. Start IQC mode
A. Start IQC Mode in the engineer mode.
B. Test the each mode.
C. If Test is success or fails, select pass or fail on pass and fail