![Huvitz HRK-9000A Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/huvitz/hrk-9000a/hrk-9000a_service-manual_2190272045.webp)
44 Auto Ref/Keratometer HRK-9000A ------------------------------------------------------------------------------------------------
button. Screen, the starting value for the PD (PD Start) in the
current encoder value is output.
3) PD-left party based in the scale and press the joystick button.
Screen, the exit value for the PD (PD End) will be printed in the
current encoder values.
4) PD distance marked on the top right of the screen if
appropriate, SAVE button and press Save. If you are not
adequate, adjust the offset of the slit spacing, and try again seen.
D. After PD Setup, engineer mode by pressing the
SAVE SYS DATA
is stored.
4-1-12. System Data Setup
A. Start System Setup in the engineer mode.
B. System Setup can be set in the type of system is as follows.
1) REF DUTY
-
Th
e brightness of a REF LED’s PWM
2) REF STEP
-
The step of a REF LED
3) ILLUM STEP
-
The step of a ILLUM(REF) LED
4) MIRE DUTY
Содержание HRK-9000A
Страница 1: ...Service Manual Auto Ref Keratometer HRK 9000A...
Страница 14: ...Auto Ref Keratometer HRK 9000A 13 3 2 Optical System...
Страница 19: ...18 Auto Ref Keratometer HRK 9000A 3 3 2 Electrical wiring 3 3 2 1 Main Board...
Страница 25: ...24 Auto Ref Keratometer HRK 9000A Objective Lens Image Lens MLA Micro Lens Array TV Camera...
Страница 27: ...26 Auto Ref Keratometer HRK 9000A Ring type light source for measuring Mire Ring TV Camera Lens for measuring...