![Huvitz HRK-9000A Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/huvitz/hrk-9000a/hrk-9000a_service-manual_2190272033.webp)
32 Auto Ref/Keratometer HRK-9000A ------------------------------------------------------------------------------------------------
4-1-4.
Check the setup data of REF (REF Setup Data)
A.
Begins
“REF DATA SETUP” in the engineer mode.
B.
Press the MODE button, and switch REF diopter mode.
C.
In selected diopter mode, cmodels of the data confirmed that the child is.
1) DIOPT
- diopter of model eye
2) GX
- average horizontal distance of REF Spots
3) GY
- average vertical spacing of REF Spots
4) SPH
- Spherical coefficients of the Zernike
5) CYL
- Cylinder coefficients of the Zernike
NOTE
(1)
If you set the REF LOW or HIGH diopter, you must set-up 0 diopter model
eye first.
(2)
In the case of whole diopter model eye is already setup, If you set-up 0D
model eye again, All of the other model eye diopter should be set up again.