Auto Blocker
3.4 Blocking axis calibration
Although the cross marker of a optic center is correct at the display, if the blocked axis is incorrect for 3 dot marked
lenses, progressive lenses or bi-focal lenses, it is needed for you to calibrate the blocking axis.
Step 1. Reset the misaligned angle 0 at “FACTORY MENU>BLOCKER”Step 2. Mark a lens with a cylindrical power
using your lensmeter.
Step 3. Go into “Calibration > BLOCKER” page.
Step 4. Put the lens to be placed at the center of the cross and paralleled along the horizontal line.
Step 4. Click “BLK CENTER” button.
Step 5. Take it out and examine the blocked angle.
Step 6. Go into “FACTORY MENU > BLOCKER” page.
Step 6. Apply the misaligned angle at the “Shift Angle”Step 7. Press the save button to apply to your calibration.
Step 8. Repeat from Step 4 to Step 7 to fit up the blocking axis.
Содержание Excelon CAB-4000
Страница 1: ......
Страница 2: ...Auto Blocker CAB 4000 1 Service Manual Auto Blocker CAB 4000...
Страница 6: ...Auto Blocker CAB 4000 5 Dimensions...
Страница 27: ...26 Auto Blocker 4 Repair Standard 4 1 Removing Side Cover Assembly P I C T U R E...
Страница 29: ...28 Auto Blocker 4 2 Removing Cover LCD Assy P I C T U R E...
Страница 31: ...30 Auto Blocker 4 3 Removing the Cover Front Assy P I C T U R E...
Страница 33: ...32 Auto Blocker 4 4 Removing the Module 1 Tracer Image Camera Inner cover P I C T U R E...
Страница 35: ...34 Auto Blocker 4 5 Removing the Module 2 Blocking Arm Lensmeter LED Projection Assy etc P I C T U R E...
Страница 37: ...36 Auto Blocker 4 6 Removing the SMPS GUI Board and Motor Board P I C T U R E...
Страница 39: ...38 Auto Blocker 4 7 Removing the Sensor within the Blocking Arm Modul P I C T U R E...
Страница 41: ...40 Auto Blocker 4 8 Removing the Blocking Theta Axis Assy...