Maintenance
771402-2 dopplex ABIlity Service Manual
56
11.5 General Inspection
The following areas should be inspected during every service:
•
Front panel keypad buttons, ensuring that they are not split or damaged
•
Cuff connectors, particularly the sealing ‘O’ rings
•
Display window & LCD
•
All areas of the enclosure
•
Mains Inlet connector for (condition & security)
•
USB socket
•
Printer paper tray
11.6 Real Time Clock Battery Check
The internal real time clock is maintained by a replaceable Lithium-ion coin cell. This battery can be expected to
last for at least five years. If time and date storage becomes unreliable, it should be replaced.
For battery replacement instructions refer to section 10
.