24
www.humancaregroup.com
seBastian II
Cleaning contact surfaces
1. The surface of the handles must be cleaned
and maintained as follows:
a)
undo the straps and remove all
parts that do not belong to the lift.
b)
clean surfaces with mild and
environmentally-friendly cleaning
agents
c)
the same applies to cleaning the
hand control.
2.
Accessible areas of the push handles and
the frame can be disinfected using mild
and environmentally-friendly detergents
3. The castors only need to be disinfected
if there are visible traces of infected or
potentially contaminated material.
Servicing
An annual service and maintenance by autho-
rized personnel is required every 12 months.
Human Care cannot be held responsible for
damage caused by willful damage, overdue
maintenance and incorrect use, or any resulting
consequential damage.
We recommend checking the product regularly
for loose parts. Regularly clean the castors.
Maintenance
During long periods of inactivity, we recom
-
mend checking all electrical and mechanical
parts once a month by carrying out a lift test
without the client. Additionally, the charging
cables should be visually checked for possible
damage after each mechanical lift or after mov
-
ing the lift to a different location.
Do not use:
• pastes, waxes, sprays,
• strong detergents, solvents and de-
tergents containing solvents, alcohol
and leather cleaners.
Using such agents can also cause hard-
ness and cracks in the material.
Any changes to the surface sheen are not
covered under the warranty!