3
Pressing all three pushbuttons simultaneously will reset the captured Peak value,the cpatured
Valley value and reset the latching relays.
Note that the state of latching relays and the Peak/Valley values are stored electronically. If
power to the unit is interrupted the latched relay states will be lost and Peak/Valley values will
be reset.
Operation During Setup
During setup programming the bar and digital display provide information to the user for
programming. The digital display is used to show 3 to 5 character alphanumeric mnemonics
for each of the setup functions and the bar display is used to provide an analog indication of
functions and function values.
Information from the user is input through pushbuttons on the front panel.
Operator and Supervisor Setup
Setup functions are divided into two separate classes: Operator and Supervisor.
Operator Setup functions are those that may need to be changed on a routine basis for normal
operation of the instrument. These functions include changing the alarm setpoint values, LED
brightness control and Lamp Test. The ability to perform alarm setpoint programming, LED
brightness control, Bar/Digital display flashing, Peak/Valley value reset and latching relay
reset may all be disallowed by disabling Operator mode in Supervisor Setup.
Supervisor functions are those that should be setup only once when the instrument is installed
and may cause damage to equipment or faulty operation if changed inadvertently. These func-
tions include calibration settings, retransmit, ID settings, bar display modes and others.
Supervisor Setup functions may be programmed only after entry of a correct password. The
password may be changed by the user in Supervisor Setup if the current password is known.
The factory default password is 0. See notes under
PASS
function regarding changing the
password.
The following pages have detailed descriptions of every function in Operator and Supervisor
Setup. For quick reference, see Section 5, Configuration Flowcharts.