36
13. Step the INCREASE ( ) pushbutton to scroll through the setup modes until you return to
R U N
.
14. Press ENTER ( ). At this point, the BarGraph will return to normal operation.
15. Recheck your BarGraph's calibration values with the lab-standard input signal source and
verify that you have accomplished the desired results. Values between each of the calibra-
tion points should be checked to insure that accuracy is maintained across the full range of
calibration values.