32 | 60
OWNERS MANUAL | Cobolt 08-01 Series | D0141-K SEPTEMBER 2020
Controlling emission in Continuous Wave operation
In this section the different ways to control the emission of each laser in continuous wave (CW) operation will
be discussed in detail. It is not recommended to use the continuous wave emission or power level controls to
turn the laser ON and OFF with high speed. See section 7 :
for instructions on high
speed emission control.
RESTART and ABORT Button
T
he ‘Restart’ button will start the complete autostart sequence including temperature stabilization and warm
up. The ‘Abort’ button cancels the autostart sequence and places the entire device in an OFF state. The key
switch is required to restart the device when in CDRH mode. The restart button corresponds to the command
‘restart’, the abort button corresponds to the system command ‘abort’.
Laser ON and Laser OFF Button
The Laser ON and Laser OFF buttons can be used to stop and start emission regardless of the operation mode
(constant power, constant current or modulation mode). Access to the Laser ON and Laser OFF buttons is
available in the summary level of the user interface as well a
s the ‘More’ window.
NOTICE
Turning the lasers ON and OFF with these buttons will require toggling of the key switch.
The ON and OFF button corresponds to the command ‘@cob1’ for
Laser ON
and ‘
l
0
’ for Laser OFF.