23 | 60
OWNERS MANUAL | Cobolt 08-01 Series | D0141-K SEPTEMBER 2020
Operation and Environmental Specifications
MLD
DPL
Power supply requirements
5 VDC, 3 A
5 VDC, 5 A
Intended use environment
Laboratory (indoor)
Maximum baseplate temperature
50
C
Ambient temperature, operation
10 - 40
º
C
Ambient temperature, storage
-10 °C to +60
º
C
Humidity
0-90 % RH non-condensing
Ambient Air pressure
950-1050 mbar
Laser Head heat sink thermal impedance at 40
C ambient
< 0.8 K/W
< 0.5 K/W
Maximum heat dissipation of Laser Head
< 12 W
< 20 W
Warm-
up time from complete “off”
< 3 min
Communication protocol
USB and RS 232
USB
Ambient temperature & pointing - 20-50 °C
< 5 µrad /
o
C
Beam position accuracy (mm)
< 0.5
Beam angle accuracy (mrad)
< 5
Electrical interfaces
All equipment connected to the system should be limited energy as described by IEC 61010:1.
Interface
Location
MLD
DPL
Input power
Laser Head
DC plug 2.5 mm / 5.5 mm female
Remote interlock connector
Laser Head
2.5 mm audio female
OEM: Molex pin 1 & 2
Data port
Laser Head
USB-type mini B
Key control Box connector
Laser Head
VGA D-SUB 15-pin male
Molex 6-pin male
Digital modulation
Laser Head
SMA female
Analog modulation
Laser Head
OEM: D-SUB 15 pin 7 & 8
OEM: Molex pin 2 & 3
Laser Head connector
Key control box
VGA D-SUB 15-pin male
Remote Interlock connector
Key control box
----
CDRH: 3.5 mm audio female
RS-232
Key control box
D-SUB 9-pin male
---
On/off modulation
Key control box
3.5 mm audio female
---
Analog modulation
Key control box
BNC female
Mechanical Interfaces
MLD
DPL
Laser Head dimensions
100 x 40 x 40 mm
Fixation holes, Laser head
= 4 x 4.2 mm; 90 mm x 30 mm
Weight
< 0.2 kg
< 0.3 kg
Mechanical Drawings