Service/maintenance
Chapter 6
OPERATION MANUAL
V2.3.0en/04.08.21//1.30
RotaCool®
41
6.2.2
Replacing temperature control hoses
Replace defective temperature control hoses
before
turning on the temperature control unit.
PROCEDURE
Drain the temperature control unit.
→
»Draining externally closed applications«
Replace defective temperature control hoses. Observe the proper disposal.
→
Page 15, section
»Proper disposal of resources and consumables«
Reconnect your external application.
→
»Connecting externally closed application«
Fill the temperature control unit with thermal fluid.
→
externally closed application«
Vent the temperature control unit.
→
»Filling and venting externally closed
Restart the temperature control unit in normal mode.
6.2.3
Clean liquefier fins (air-cooled temperature control unit)
Manual cleaning
RISK OF BEING CUT ON THE LIQUEFIER FINS
Wear suitable cut-resistant gloves for cleaning work.
Depending on the ambient conditions, use cleaning equipment such as vacuum cleaners and/or
a hand brush/brush. Follow the local regulations when cleaning. Do not clean the liquefier fins
in a clean room with items like a brush and do not use a vacuum cleaner without an extra-fine
particle filter.
Cleaning using pointed or sharp-edged tools
DAMAGE TO THE LIQUEFIER FINS
Clean the liquefier fins using suitable cleaning appliances.
Make sure there is adequate ventilation (removal of waste heat, fresh air supply) for the tempera-
ture control unit, in case of
air cooling, maintain wall clearance
.
→
Page 19, section
illustrations of the cooling variants«
and
→
The liquefier fins must be cleaned (dust) from time to time as only then will the temperature con-
trol unit perform at its maximum cooling capacity.
The grille is located at the bottom of your temperature control unit.
PROCEDURE
Ventilation grille on the underside (table-top units)
Cleaning of liquefier fins at the underside when the temperature control unit is filled
DAMAGE CAUSED BY THERMAL FLUID PENETRATING THE TEMPERATURE CONTROL UNIT
Empty the temperature control unit before cleaning the liquefier fins at the underside of the
temperature control unit.
Switch off the temperature control unit.
Disconnect the temperature control unit from the power supply.
Drain the thermal fluid from the temperature control unit.
→
Tilt the temperature control unit to remove the grille (if available) in front of the liquefier fins.
Clean the liquefier fins using suitable cleaning appliances. Observe the local regulations and
ambient conditions when selecting cleaning appliances.
Make sure the liquefier fins are not damaged or deformed as this will impair the air flow.
Re-mount the ventilation grille after cleaning work.
Connect the temperature control unit to the power supply.
Refill the temperature control unit with thermal fluid.
→
externally closed application«