I-V400w - I-V500w - SOLAR I-Vw - SOLAR I-Ve
EN - 44
23. The instrument displays:
Vdc = DC output voltage from module/string
Irr = irradiance measured by supplied reference cell
Tc = temperature of module. The herewith options are
available for this field:
AUTO
automatic mode of temperature measurement
Number
MAN or AUX mode of temp. measurement
“ - - - “
AUX mode with no connected probe
Module = type of selected module
Temp = measurement mode of module temperature
Start = measurement activation mode
The symbol “
”
fixed
at display means that a stable
RF connection is established with the remote unit
SOLAR-02
15/05/10 15:34:26
V d c
=
3 6 7
V
I r r
=
1 0 4 5
W / m 2
T c
=
4 5
° C
Module: SUNPOWER 210
Temp: Auto
Start: Manual
Select
I – V
24. Press
ENTER
key, select the “
Activate Rec.
” item and
confirm with
ENTER
. The message “Remote unit in
recording…” is displayed on the meter. In this condition is
possible to start the I-V curve tests always considering
that the complete results will be available only after the
stop of recording of SOLAR-02 and transfer them from the
same remote unit
15/05/10 15:34:26
V d c
=
0 . 0
V
I r r
= - - - W / m 2
T c
= - - - ° C
A c t i v a t e R e c .
S e t t i n g s
M e a s . T y p e
Select
I – V
CAUTION
At the press of
GO/STOP
key some different error messages can be
displayed by the meter (see § 6.4) and, due of this, the test cannot start.
Check and remove, if possible, this error message before start the test
25. Press
GO/STOP
key (in Manual activation mode) to start
the test. If none of the previous error messages are
detected, the message “
Measuring…
” is shown on the
display for some seconds, depending on the level of
power during test.
15/05/10 15:34:26
V d c
=
3 6 7
V
I r r
=
- - -
W / m 2
Tc
=
Auto
°C
Module: SUNPOWER 210
Measuring…
Select
I - V