5. OPERATING INSTRUCTIONS 

5.1.  MEASUREMENT OF AIR SPEED AND TEMPERATURE 

 

 

CAUTION 

 

Do not touch not expose the internal sensors found on the top side of the 
telescopic probe to mechanical shocks, in order to prevent damaging them. 

 

 

Fig. 7: Measurement of air speed and temperature 

 

1.  Insert the external probe on the top side by means of the relevant terminal (see Fig. 1 

– 

part 1). Pay attention to the arrow found on the connector of the probe (see Fig. 7),  
which must be directed towards the front side of the instrument 

2.  Switch on the instrument using key 

ON/OFF

3. Press key 

UNIT

 to set the measuring unit of air speed (see § 4.3.5) and key 

UNIT

 

to set the measuring unit of air temperature (see § 4.3.6) on the secondary display after 
press key 

FLOW/TEMP

 

4.  Let the protection slide (see Fig. 1 

– 

part 15) until the sensors are covered and check 

that the indication on the main display is 

0.00

. In case it is not, please carry out the 

zeroing procedure (see § 4.3.2) 

5.  If necessary, expand the telescopic probe and position the air speed sensor in a 

parallel direction to the air flow as indicated by the arrow found on the top of the probe 
itself (see Fig. 7). 

6.  The value of air speed is indicated on the main display, while air temperature is shown 

on the secondary display. 

7. Press key 

HLD

 to freeze the reading on the display. 

8.  To use MAX, MIN functions, see § 4.3.8. 
 

Содержание HTA105

Страница 1: ...Copyright HT ITALIA 2018 Version EN 1 00 19 03 2018 ENGLISH User manual...

Страница 2: ...4 3 8 Key MAXMIN 5 4 3 9 Key MEAN 5 4 4 Instrument settings 6 5 OPERATING INSTRUCTIONS 8 5 1 Measurement of air speed and temperature 8 5 2 Measuring volumetric flow rate and air humidity 9 5 3 Calcu...

Страница 3: ...ng observe the instructions given in this manual improper use could damage the instrument or its components 1 1 PRELIMINARY INSTRUCTIONS We recommend following the normal safety rules devised to prote...

Страница 4: ...ked from an electric as well as mechanical point of view All possible precautions have been taken so that the instrument is delivered undamaged However we recommend generally checking the instrument i...

Страница 5: ...CRIPTION OF THE DISPLAY CAPTION 1 Speed measurement icon 2 Active Settings 3 Active Data HOLD 4 Indication of minutes seconds 5 Indication of active Auto Power OFF 6 Low battery symbol 7 Main display...

Страница 6: ...allows selecting the measuring unit of air temperature shown on the secondary display between the options C Celsius and F Fahrenheit selecting the measuring unit of air volumetric flow rate between t...

Страница 7: ...n the display 4 Press keys UNIT or UNIT to select the desired measuring unit among the options in2 m2 or ft2 5 Press key ENT to confirm The screen in Fig 4 left side is shown on the display Fig 4 Sett...

Страница 8: ...PA mbar not measured by the instrument is displayed Fig 6 Setting of atmospheric pressure value CAUTION The atmospheric pressure value is important for the necessary compensation in order to obtain ac...

Страница 9: ...y UNIT to set the measuring unit of air speed see 4 3 5 and key UNIT to set the measuring unit of air temperature see 4 3 6 on the secondary display after press key FLOW TEMP 4 Let the protection slid...

Страница 10: ...shown on the display 6 Press key UNIT to select measuring unit CMM or CFM on the secondary display see 4 3 6 7 Let the protection slide see Fig 1 part 15 until the sensors are covered and check that...

Страница 11: ...to be carried out with the same procedure as for air speed 5 Press key MEAN Symbols and Mean are shown in the bottom part of the display while the number of the measuring spots are show on the main d...

Страница 12: ...as for air speed 5 Press and hold 2s key MEAN Symbols Mean and and air speed are shown respectively in the bottom part of the display and on the secondary display while symbols 0000 and mm ss timer ex...

Страница 13: ...ation make sure you have removed the probe from the input terminal When symbol is shown on the display air speed indication is flashing and in these conditions the measured value does not comply with...

Страница 14: ...122 0 F 0 1 F 1 8 F Measurement of air relative humidity Range Resolution Accuracy 0 RH 100 RH 0 1 RH 5 RH 7 1 1 General characteristics Mechanical characteristics Size L x W x H 190 x 65 x 45mm 7 x 3...

Страница 15: ...e humidity 80 RH Storage temperature 10 C 60 C 14 F 140 F Storage humidity 80 RH Max operating altitude 2000m 6562ft This instrument complies with European Directive EMC 2014 30 EU This instrument sat...

Страница 16: ...consequence of improper packaging Repairs which may become necessary as a consequence of interventions performed by unauthorized personnel Modifications to the instrument performed without the manufa...

Отзывы: