3-10
Chapter 3
Gauging and Making Connections
General Cleaning Procedures
especially when cleaning a female connector (to avoid snagging the
swab on the center conductor contact fingers).
Drying the Connector
After cleaning, blow the connector dry with a gentle stream of clean
compressed air or nitrogen. Always completely dry a connector before
you reassemble or use it.
Inspecting Each Connector Again
Inspect the connector again under a magnifying glass to be sure that no
particles or residues remain.