![HP E3494A Скачать руководство пользователя страница 27](http://html.mh-extra.com/html/hp/e3494a/e3494a_installation-and-service-manual_161547027.webp)
performance, increased LAN traffic, and even failure to power up (if the
BOOTP server becomes inactive).
7
Verify your HP processor probe is now active and on the network.
See "To verify LAN communications" on page 20.
To set the 10BASE-T configuration switches
Set switches S7 and S8 to CLOSED unless one of the following conditions is
true:
•
If the LAN cable exceeds the standard length, set switch S7 to OPEN.
The HP processor probe has a switch-selectable, twisted-pair receiver
threshold. With switch S7 set to OPEN, the twisted-pair receiver threshold is
lowered by 4.5 dB. This should allow you to use cable lengths of up to about
200 meters. If you use a long cable, you should consult with your LAN
cabling installer to ensure that:
•
The device at the other end of the cable has long cable capability, and
•
The cable is high-grade, low-crosstalk cable with crosstalk attenuation
of greater than 27.5 dB.
When switch S7 is set to CLOSED, the LAN port operates at standard
10BASE-T levels. A maximum of 100 meters of UTP cable can be used.
•
If your network doesn’t support Link Beat integrity checking or if the
HP processor probe is connected to a non 10BASE-T network (such
as StarLAN) set this switch to LINK BEAT OFF (0 or OPEN).
In normal mode (switch S8 set to CLOSED), a link integrity pulse is
transmitted every 15 milliseconds in the absence of transmitted data. It
expects to receive a similar pulse from the remote MAU. This is the standard
link integrity test for 10BASE-T networks. If your network doesn’t support
the Link Beat integrity checking or if the processor probe is used on a non
10BASE-T network (such as StarLAN) set this switch to LINK BEAT OFF
(OPEN).
N o t e
Setting switch S8 to OPEN when Link Beat integrity checking is required by
your network will cause the remote MAU to disable communications.
Connecting to the Host Computer
To set the 10BASE-T configuration switches
19
Содержание E3494A
Страница 3: ...iii ...
Страница 4: ...iv ...
Страница 8: ...Contents viii ...
Страница 9: ...Part 1 Installation Installation 1 ...
Страница 10: ...Installation 2 ...
Страница 11: ...1 Connecting to the Host Computer Connecting to the Host Computer 3 ...
Страница 34: ...Connecting to the Host Computer To verify serial communications 26 ...
Страница 35: ...2 Connecting to the Target System Connecting to the Target System 27 ...
Страница 38: ...Connecting to the Target System To connect to a target system via the IEEE 1149 1 JTAG Port 30 ...
Страница 39: ...Part 2 Using the HP processor probe Using the HP processor probe 31 ...
Страница 40: ...Using the HP processor probe 32 ...
Страница 62: ...Configuring the HP Processor Probe for PowerPC 603 603e Processor Memory Cache Configuration Items 54 ...
Страница 63: ...Part 3 Reference Reference 55 ...
Страница 64: ...Reference 56 ...
Страница 65: ...4 Designing a Target System Designing a Target System 57 ...
Страница 72: ...Designing a Target System PowerPC 603 JTAG Interface Connections and Resistors 64 ...
Страница 73: ...5 Specifications and Characteristics Specifications and Characteristics 65 ...
Страница 77: ...6 Updating Firmware Updating Firmware 69 ...
Страница 83: ...7 Solving Problems Solving Problems 75 ...
Страница 102: ...Solving Problems To obtain replacement cables 94 ...
Страница 104: ...U Ultra target board 61 update firmware 70 USER light 77 V version firmware 73 W workstation files 25 Index 96 ...