![HP E3491A Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/hp/e3491a/e3491a_installation-and-service-manual_161546056.webp)
Problems with the HP E3491A Pentium Probe
Itself
To run the power up self test
1
Unplug the HP E3491A Pentium Probe, then plug it in.
2
Watch the status lights. They should show the following pattern:
❍
= LED is off
●
= LED is on
✳
= Not applicable (LED is off or on)
Normal sequence during power up self test
Pwr/Target
LEDs
Meaning
1
●●●●
Initial power up, system reset
2
●❍❍❍
XILINX array initialized successfully
3
●❍❍●
XILINX array tested successfully
4
●❍●❍
BOOT ROM space tested successfully
5
●❍❍●
GENERIC ROM space tested successfully
6
●❍●❍
DRIVER ROM space tested successfully
7
●❍❍●
RESERVED ROM space tested successfully
8
●❍●❍
RAM tested successfully
9
●❍●●
LAN internal feedback tested successfully
10
●❍●●
Start system, load drivers, initialize LAN
If the power up self test fails, the RESET LED will flash the number of the
test, then stay lit.
If any of the LEDs fail to change, or all of them remain on, there is a system
failure.
Following power up, the LEDs will enter one of the following states:
Solving Problems
To run the power up self test
48
Содержание E3491A
Страница 3: ...HP E3491A Pentium Probe Figure 1 iii ...
Страница 4: ...iv ...
Страница 8: ...Contents viii ...
Страница 9: ...Part 1 Installation ...
Страница 10: ...2 ...
Страница 11: ...1 Connecting to the HP 16505A Prototype Analyzer ...
Страница 29: ...2 Connecting to the Target System ...
Страница 34: ...26 ...
Страница 35: ...Part 2 Reference 27 ...
Страница 36: ...28 ...
Страница 37: ...3 Designing a Target System ...
Страница 40: ...32 ...
Страница 41: ...4 Specifications and Characteristics 33 ...
Страница 46: ...38 ...
Страница 47: ...5 Solving Problems Solving Problems 39 ...
Страница 66: ...58 ...