90 CLEAR SCREEN
100 PRINT "Install Component Assembly and Test Cable"
110 PRINT
120 PRINT " 1. Turn mainframe power OFF"
130 PRINT " 2. Install HP E1330A/B component assembly into
mainframe "
140 PRINT " 3. Attach test cable to component assembly"
150 PRINT " 4. Turn mainframe power ON "
160 PRINT " 5. Press Continue when ready to begin testing "
170 PAUSE
180 CLEAR SCREEN
190 OUTPUT @Dio;"*IDN?"
200 ENTER @Dio;A$
210 IF A$[17,21]<>"E1330" THEN
220 PRINT "Incorrect Card Type detected"
230 STOP
240 END IF
250 PRINT "Bit walk test of all ports for read and write (no handshake)"
260 FOR I=0 TO 7
270 FOR J=0 TO 3
280 OUTPUT @Dio;"SOUR:DIG:DATA"&VAL$(J)&"
#B"&Bit_walk$(I)
290 IF J<2 THEN
300 OUTPUT @Dio;"MEAS:DIG:DATA"&VAL$(J+2)&"?"
310 ELSE
320 OUTPUT @Dio;"MEAS:DIG:DATA"&VAL$(J-2)&"?"
330 END IF
340 ENTER @Dio;A$
350 Number=VAL(A$)
360 B$=DVAL$(Number,2)
370 IF B$[25,32]<>Bit_walk$(I) THEN
380 Fail = 1
390 PRINT "Failure of bit walk"
400 PRINT "WRITE at port ";J
410 IF J<2 THEN
420 PRINT "READ at port ";J+2
430 ELSE
440 PRINT "READ at port ";J-2
450 END IF
460 END IF
470 NEXT J
480 NEXT I
490 PRINT "End of bit walk test"
22 Verification Tests
Chapter 2