Functional
Verification
Test
The Functional Verification Test for the HP E1330A/B modules consists of
sending the *IDN? command and checking the response. This test can be
used to verify that the module is connected properly and is responding to a
basic command.
Procedure
1.
Verify that the module is properly installed in mainframe
2.
Verify that the mainframe has passed its power-on test.
3.
Send *IDN? to the module (see example following)
4.
The return should be as follows (revision number may vary):
HEWLETT-PACKARD,E1330A,0,A.06.00
NOTES
If the primary address setting, secondary address setting, or the interface
select code is set incorrectly, the module will not respond. Verify proper
address selection before troubleshooting.
Both the HP E1330A and HP E1330B return the string shown in Step 4. The
HP E1330B will return "E1330A" in response to the *IDN? query.
Example
An example follows which uses an HP 9000 Series 300 computer with HP
BASIC and a module address of 70918.
10 DIM A$[100]
20 OUTPUT 70918;"*IDN?"
Send the ID command
30 ENTER 70918;A$
Get response
40 PRINT A$
50 END
Operation
Verification
Test
The procedures in this section are used to provide a high level of confidence
that the module is meeting published specifications. The Operation
Verification test is a subset of the Performance Verification tests and is
suitable for checkout after performing repairs.
The Operation Verification Test is performed by completing the Digital
Test (Test 2-1) as described in the Performance Verification test procedures.
This test is usually sufficient to verify that the module is meeting its
specifications.
18 Verification Tests