HP D300 Digital Dispenser
23
b.
Lift a new Cassette out of the clamshell container and place it white logo side up into the
cutout on the Dispenser deck.
c.
Make sure it is fully inserted so the top surface of the Cassette is the same height as the
top of the deck all the way around the cassette.
Figure 4-21
Loading a new Dispensehead Cassette
3.
Once the Cassette is fully inserted, click
OK
in the Load Cassette dialog box (see
). The pogo mechanism will engage and the Dispenseheads will be electrically tested to
ensure quality. Then the Load Fluid dialog box appears. See
4-5-4 Load fluid into the Dispensehead
When the Load Fluid dialog box opens, the OK and Skip buttons flash above the first Dispensehead
that will be loaded (see
), and a blue LED illuminates the first Dispensehead on the front
of the pogo mechanism.
1.
Fill a low volume pipette with at least the amount of fluid that is indicated in the Load Fluid
dialog box, or fill a multipipetter with the appropriate number of fluids in the proper order as
shown in the Load Fluids dialog box. If a multipipetter is used, use the Max. Load Volume for
all Dispenseheads as shown on the right side of the dialog, as circled in
. The
source plate and source well location of the fluid to load will also be identified in the Load Fluid
dialog box if that information was entered in the Fluids List. Be sure to load the proper
compound into the proper Dispensehead.
CAUTION:
Using a multichannel pipette enables the loading of multiple Dispenseheads
simultaneously. Make sure that you fill the Dispenseheads with the
LARGEST
load volume of the
set of Dispenseheads you are loading. It is indicated in the upper right of the dialog box
(―Max.
Содержание Compaq d300 Series
Страница 1: ...HP D300 Digital Dispenser Operating Manual ...
Страница 72: ...HP D300 Digital Dispenser 72 Appendix C Quick Start Guide ...