81
Chapter 2, Troubleshooting
More About Step 1- Run the Power-Up Diagnostic
N:\mkt\MANUALS\HP8924C\ALR\trouble.fb
More About Step 2- Run the Functional Diagnostics
USER Keys
The numbered control fields in the upper-right corner of the TESTS screens
correspond to the USER keys to the right of the screen.
Frequently Encountered Error Messages
Error messages that appear on the second line of the Test Set’s display frequently
occur while the Functional Diagnostics are running. The most complete and
general list of error messages is in the Error Messages chapter of the Test Set’s
User’s Guide . Some messages relating specifically to troubleshooting can be
found in the Error Messages chapter of this manual. Some of the messages you
can expect to occur while running the Functional Diagnostics are as follows:
Functional Diagnostics measurements commonly generate the message
”Direct latch write occurred. Cycle power when done
servicing.”
The message appears the first time the diagnostic program directly
addresses a latch. The message should be ignored until you wish to make a normal
(not a diagnostic) measurement with the Test Set. To clear this message the Test
Set should be turned off and back on again.
The message
”Printer does not respond.”
usually indicates that one or
more settings on the TESTS (Printer Setup) screen are set wrong for your printer.
Also, check that the printer’s power is on and that it is correctly cabled. For HP-IB
printers make sure the printer is correctly addressed. If a serial printer is used, you
may have to change the serial communication settings on the I/O CONFIGURE
screen (press SHIFT then PREV to get to this screen). The message times-out
after a few seconds, and the output destination is changed to CRT by the program.
Some error messages you might encounter when running the Functional
Diagnostics are the following:
”ERROR 173 IN XXXX Active/system controller req’d”
(where
”XXXX”
represents a line number) indicates that the Test Set’s internal IBASIC
computer must be set as a system controller for some reason. This usually
indicates that the
Printer Port
field of the TESTS (Printer Setup) screen was
set to
HP-IB
but the
Mode
field on the I/O CONFIGURE screen is set to
Talk&Lstn
instead of
Control
. Change the mode setting to
Control
and run
the diagnostic again.
Содержание 8924E
Страница 52: ...Contents 52 N mkt MANUALS HP8924C ALR 8924CE TOC A Error Messages ...
Страница 53: ...Contents 53 N mkt MANUALS HP8924C ALR 8924CE TOC Index 633 ...
Страница 54: ...Contents 54 N mkt MANUALS HP8924C ALR 8924CE TOC ...
Страница 70: ...70 Chapter 1 Introduction Hardware and Firmware Enhancements N mkt MANUALS HP8924C ALR intro fb ...
Страница 137: ...137 N mkt MANUALS HP8924C ALR per_adj fb 4 Periodic Adjustments ...
Страница 448: ...448 Chapter 9 HP 8924C E Specifications General Specifications N mkt MANUALS HP8924C ALR specs fb ...
Страница 449: ...449 N mkt MANUALS HP8924C ALR service fb 10 Service Screen ...
Страница 454: ...454 Chapter 10 Service Screen Troubleshooting with the Service Screen N mkt MANUALS HP8924C ALR service fb ...
Страница 455: ...455 N mkt MANUALS HP8924C ALR blkdia fb 11 Block Diagrams ...
Страница 522: ...522 Chapter 11 Block Diagrams Assembly Details N mkt MANUALS HP8924C ALR blkdia fb ...
Страница 530: ...530 Chapter 12 HP 8924C Replaceable Parts Replaceable Parts Listing N mkt MANUALS HP8924C ALR 24c_rpl fb ...
Страница 538: ...538 Chapter 13 HP 8924E Replaceable Parts Replaceable Parts Listing N mkt MANUALS HP8924C ALR 24e_rpl fb ...
Страница 613: ...613 N mkt MANUALS HP8924C ALR errmssg fb A Error Messages ...
Страница 640: ...Index 640 N mkt MANUALS HP8924C ALR 8924CE IX V VOLUME control 61 W Write Protect Cal Data 143 ...