76
Chapter 2, Troubleshooting
How to Troubleshoot the Test Set
N:\mkt\MANUALS\HP8924C\ALR\trouble.fb
is configured later in this procedure
7. Under
Output Results For:
•
Select
All
to display (and print) all test information.
•
Select
Failures
to only display (and print) failed test points.
8. Under
If Unit-Under-Test Fails:
•
Select
Continue
to continue to the next test point.
•
Select
Stop
to pause testing at that point.
9. Under
Test Procedure Run Mode:
•
Select
Continuous
to run the tests continuously.
•
Select
Single Step
to pause after each measurement.
10. Under
Autostart Test Procedure on Power-Up:,
verify that the setting
is
Off
.
CONFIGURE A
PRINTER
Only perform this next step if you want to print test results to a printer, otherwise
go to step 16.
11. Press the TESTS key to return to the TESTS (Main Menu) screen.
12. Under
SET UP TEST SET:
, select
Printer Setup
to access the TESTS (Printer
Setup) screen.
13. Under
PRINT SETUP:
, select the
Model:
of printer to use.
14. Set the
Printer Port:
for the rear-panel connector your printer is connected to.
If an HP-IB printer is used, you need to enter the printer’s two-digit bus address when
the
Printer Adrs
field appears (Example; enter
01
for bus address 701). Also,
press the SHIFT key, then the PREV key to access the I\O CONFIGURE screen, and
set the
Mode
field to
Control
.
15. Under
PAGE CONTROL:
, set the
Lines/Page
: and Form Feed (
FF at Start:
,
and
FF at End:
) parameters if necessary.
Begin Testing
16. Select the
Run Test
field (or press K1) and wait for the “
Loading program -
Enter SHIFT-CANCEL to abort.
” message to go away.
17. Choose the test or tests to run by selecting
Up
or
Down
to move the pointer and then
choosing
Select
. When running the Audio, RF, or CDMA diagnostics for the first
time, you should run “
All xxx Tests
” (xxx=Audio, RF, or CDMA).
18. Follow the instructions on the screen.
19. As the tests run, you can alter test execution conditions by selecting:
•
Loop
to run the test continuously.
•
Pause
to pause the tests.
•
Stp Fail
(stop-on-failure) to pause the tests when a failure is detected.
•
Sgl Step
(single-step) to pause the test after each measurement.
Содержание 8924E
Страница 52: ...Contents 52 N mkt MANUALS HP8924C ALR 8924CE TOC A Error Messages ...
Страница 53: ...Contents 53 N mkt MANUALS HP8924C ALR 8924CE TOC Index 633 ...
Страница 54: ...Contents 54 N mkt MANUALS HP8924C ALR 8924CE TOC ...
Страница 70: ...70 Chapter 1 Introduction Hardware and Firmware Enhancements N mkt MANUALS HP8924C ALR intro fb ...
Страница 137: ...137 N mkt MANUALS HP8924C ALR per_adj fb 4 Periodic Adjustments ...
Страница 448: ...448 Chapter 9 HP 8924C E Specifications General Specifications N mkt MANUALS HP8924C ALR specs fb ...
Страница 449: ...449 N mkt MANUALS HP8924C ALR service fb 10 Service Screen ...
Страница 454: ...454 Chapter 10 Service Screen Troubleshooting with the Service Screen N mkt MANUALS HP8924C ALR service fb ...
Страница 455: ...455 N mkt MANUALS HP8924C ALR blkdia fb 11 Block Diagrams ...
Страница 522: ...522 Chapter 11 Block Diagrams Assembly Details N mkt MANUALS HP8924C ALR blkdia fb ...
Страница 530: ...530 Chapter 12 HP 8924C Replaceable Parts Replaceable Parts Listing N mkt MANUALS HP8924C ALR 24c_rpl fb ...
Страница 538: ...538 Chapter 13 HP 8924E Replaceable Parts Replaceable Parts Listing N mkt MANUALS HP8924C ALR 24e_rpl fb ...
Страница 613: ...613 N mkt MANUALS HP8924C ALR errmssg fb A Error Messages ...
Страница 640: ...Index 640 N mkt MANUALS HP8924C ALR 8924CE IX V VOLUME control 61 W Write Protect Cal Data 143 ...