4. IF Amplitude Adjustments
3. Set the spectrum analyzer controls as follows:
Center frequency
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.7 MHz
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
200kHz
Reference level
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
- 6 0
Attenuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
dB/division
dB/DIV
Resolution bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
300
Videobandwidth
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
100 Hz
4. On the spectrum analyzer, press
and IF ADJ ON OFF so OFF is underlined.
5. Set the HP 3335A controls as follows:
Frequency
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -55dBm
6. Note the marker value. Ideally it should read -60
fO.l
7. If the marker reads below -60.1
rotate
CAL OSC AMPTD one-third
turn clockwise for every 0.1
below -60
If the marker reads above -59.9
rotate
CAL OSC AMPTD one-third turn counter clockwise for every 0.1
above -60
See Figure 2-8 for the location of
A change in the displayed
amplitude will not be seen until ADJ
STATE is pressed.
If
has inadequate range, refer to “Inadequate CAL OSC AMPTD
Range” in Chapter 9, “IF Section.”
8. Press
IF STATE . After allowing the analyzer time to complete the
adjustments, the displayed amplitude and marker reading should change.
9. Repeat steps 7 and 8 until the marker reads -60
fO.l
10. Disconnect the test cable from
and reconnect W29 to
11. Set the spectrum analyzer reference level to -60
If markers are displayed, press
and
OFF .
12. Set the HP 3335A
to -60
13. Connect a BNC cable between the 50 output of the HP 3335A and the spectrum
analyzer INPUT
14. On the spectrum analyzer, press
and
. Use the front-panel knob or
step keys to place the peak of the displayed signal 3
to 5
below the reference level.
Press
(
PEAK SEARCH
)
and MARKER DELTA on the spectrum analyzer. Set the spectrum
analyzer reference level to -10
16. Change the HP 3335A
to -10
17. Press
on the spectrum analyzer.
2-28 Adjustment Procedures
Содержание 8562E
Страница 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Страница 158: ...A9 Input Attenuator 9 c 9 d 1 slll3e Figure 4 15 A9 Mounting Screw at Right Side Frame Assembly Replacement 4 33 ...
Страница 172: ...Procedure 13 Rear Frame Rear Dress Panel Figure 4 23 A6 Power Supply Cover SP14E Assembly Replacement 4 47 ...
Страница 204: ...I _I 8560E Series I I sl134e F i g u r e 5 2 P a r t s I d e n t i f i c a t i o n C o v e r A s s e m b l y I I ...
Страница 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Страница 207: ...8 5 6 2 E sl128e 12 9 p3 CES F i g u r e 5 5 P a r t s I d e n t i f i c a t i o n F r o n t F r a m e ...
Страница 213: ...w51 W2 W52 N5A A4 _ w 5 5 w27 Figure 6 3 Top View A2 and A3 Unfolded S K 1 5 7 Major Assembly and Cable Locations 6 5 ...
Страница 217: ...w20 C O A X 6 w 7 w 5 9 C O A X a 3 9 sj138e Figure 6 7 Al6 Fast ADC Option 007 Major Assembly and Cable Locations 6 9 ...
Страница 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Страница 271: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......
Страница 314: ......
Страница 355: ......
Страница 356: ......
Страница 357: ......
Страница 374: ......
Страница 416: ... i j i I i i j i i i i i i i i I i i in Figure 11 9 HP 6562E Sweep and Scan Ramps 11 42 Synthesizer Section ...
Страница 425: ......
Страница 426: ......
Страница 450: ......
Страница 469: ......
Страница 470: ......
Страница 474: ......