Procedure 7.
High Voltage Assembly
Procedure 7.
High Voltage Assembly
The A6 power supply and
high voltage assemblies contain lethal voltages
with lethal currents in all areas. Use extreme care when servicing these
assemblies. Always disconnect the power cord from the instrument before
beginning this replacement procedure. Failure to follow this precaution can
represent a shock hazard which may result in personal injury.
1. Disconnect the power cord from the spectrum analyzer.
2. Remove the spectrum analyzer cover assembly as described in “Procedure 1. Spectrum
Analyzer Cover.”
3. Fold out the A2, A3, A4, and A5 assemblies as described in “Procedure 5. A2, A3, A4,
and A5 Assemblies.”
4. Place the spectrum analyzer top-side-up on the work bench.
The voltage potential at
is
Disconnect at the CRT with
Failure to properly discharge
may result in severe
electrical shock
to
personnel and damage to the instrument.
5. Connect the spectrum analyzer line-power cord to provide proper grounding while
discharging the
post-accelerator cable. Make sure that the spectrum analyzer
line-power switch is in the off position.
6. Connect a high voltage probe
such as the HP 34111A to a voltmeter with a
10
input.
7. Connect the clip lead of the probe (ground) to the chassis of the spectrum analyzer.
8. Slip the tip of the high voltage probe under the rubber shroud of the
post-accelerator cable to obtain a reading on the voltmeter. See Figure 4-2.
9. Keep the high voltage probe on the post-accelerator connector until the voltage has
dropped to a voltmeter reading of less than 5
(less than 5 V at the connector). This
normally takes about 30 seconds.
10. Disconnect the line-power cord from the spectrum analyzer.
11. Using a small screwdriver with the shank in contact with the CRT shield assembly, slip
the tip of the screwdriver under the rubber shroud of the
post-accelerator cable
and short the cable to ground on the CRT shield assembly. See Figure 4-2.
12. Pry out the black grommet protecting post-accelerator cable
from the CRT
shield assembly.
13. Carefully unsnap the
post-accelerator cable from the CRT and discharge it by
shorting the cable to chassis ground on the CRT shield assembly.
14. Remove the three screws securing the power supply shield to the power supply and remove
the shield.
Assembly Replacement 4-25
Содержание 8562E
Страница 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Страница 158: ...A9 Input Attenuator 9 c 9 d 1 slll3e Figure 4 15 A9 Mounting Screw at Right Side Frame Assembly Replacement 4 33 ...
Страница 172: ...Procedure 13 Rear Frame Rear Dress Panel Figure 4 23 A6 Power Supply Cover SP14E Assembly Replacement 4 47 ...
Страница 204: ...I _I 8560E Series I I sl134e F i g u r e 5 2 P a r t s I d e n t i f i c a t i o n C o v e r A s s e m b l y I I ...
Страница 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Страница 207: ...8 5 6 2 E sl128e 12 9 p3 CES F i g u r e 5 5 P a r t s I d e n t i f i c a t i o n F r o n t F r a m e ...
Страница 213: ...w51 W2 W52 N5A A4 _ w 5 5 w27 Figure 6 3 Top View A2 and A3 Unfolded S K 1 5 7 Major Assembly and Cable Locations 6 5 ...
Страница 217: ...w20 C O A X 6 w 7 w 5 9 C O A X a 3 9 sj138e Figure 6 7 Al6 Fast ADC Option 007 Major Assembly and Cable Locations 6 9 ...
Страница 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Страница 271: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......
Страница 314: ......
Страница 355: ......
Страница 356: ......
Страница 357: ......
Страница 374: ......
Страница 416: ... i j i I i i j i i i i i i i i I i i in Figure 11 9 HP 6562E Sweep and Scan Ramps 11 42 Synthesizer Section ...
Страница 425: ......
Страница 426: ......
Страница 450: ......
Страница 469: ......
Страница 470: ......
Страница 474: ......