Unable to load CONDITIONS file from listed system file location
Program attempted to load the CONDITIONS file from the listed system mass storage file
location. Check the msus of the System mass storage file location and the presence of the
CONDITIONS file.
Unable to load data from <sensor filename>
Program found power sensor data file, but could not read the data from the file. Use the
Sensor Utilities to delete the file and enter new data.
Unable to obtain catalog from <system mass storage file location>
The system could not verify that the system mass storage file location entered was available.
Check the msus of the system mass storage file location.
Unable to reach power level of <value>
Program was unable to set the source amplitude for a desired power meter reading. Check the
test setup.
WARNING: CONDITIONS file from system file location is not compatible!
CONDITIONS files from Operation Verification programs for the HP 8560 A/B-Series, or
HP 8560 E-Series analyzers are not compatible with the frequency response adjustment
software. Put the frequency response adjustment software in a different directory than the
Operation Verification programs, if possible. Default CONDITIONS data will be used.
Change the mass storage file location to a different location than either of the operation
verification programs. Then, modify the conditions data as necessary and press STORE
.
6481A Sensor cal data minimum frequency not
50 MHz
Program requires the HP 8481A power sensor to have a Cal Factor at or below 50 MHz. Use
Sensor Utilities to add a Cal Factor at or below 50 MHz.
8481A Sensor cal data maximum frequency not
300 MHz
Program requires the HP 8481A power sensor to have a Cal Factor at or above 300 MHz. Use
Sensor Utilities to add a Cal Factor at or above 300 MHz.
8485A Sensor cal data minimum frequency not = 10 MHz
Program requires the HP 8485A power sensor to have a Cal Factor at 10 MHz. Use Sensor
Utilities to add a Cal Factor at 10 MHz.
8485A Sensor cal data maximum frequency not
26.5
Program requires the HP 8485A power sensor to have a Cal Factor at or above 26.5
Use
Sensor Utilities to add a Cal Factor at or above 26.5
3-18
Frequency Response Adjustment Software
Содержание 8562E
Страница 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Страница 158: ...A9 Input Attenuator 9 c 9 d 1 slll3e Figure 4 15 A9 Mounting Screw at Right Side Frame Assembly Replacement 4 33 ...
Страница 172: ...Procedure 13 Rear Frame Rear Dress Panel Figure 4 23 A6 Power Supply Cover SP14E Assembly Replacement 4 47 ...
Страница 204: ...I _I 8560E Series I I sl134e F i g u r e 5 2 P a r t s I d e n t i f i c a t i o n C o v e r A s s e m b l y I I ...
Страница 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Страница 207: ...8 5 6 2 E sl128e 12 9 p3 CES F i g u r e 5 5 P a r t s I d e n t i f i c a t i o n F r o n t F r a m e ...
Страница 213: ...w51 W2 W52 N5A A4 _ w 5 5 w27 Figure 6 3 Top View A2 and A3 Unfolded S K 1 5 7 Major Assembly and Cable Locations 6 5 ...
Страница 217: ...w20 C O A X 6 w 7 w 5 9 C O A X a 3 9 sj138e Figure 6 7 Al6 Fast ADC Option 007 Major Assembly and Cable Locations 6 9 ...
Страница 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Страница 271: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......
Страница 314: ......
Страница 355: ......
Страница 356: ......
Страница 357: ......
Страница 374: ......
Страница 416: ... i j i I i i j i i i i i i i i I i i in Figure 11 9 HP 6562E Sweep and Scan Ramps 11 42 Synthesizer Section ...
Страница 425: ......
Страница 426: ......
Страница 450: ......
Страница 469: ......
Страница 470: ......
Страница 474: ......