HP 85027A/B/C Installation
INTRODUCTION
SECTION 2
INSTALLATION
This section provides information
about initial inspection, preparation for use, mating
connectors, packaging, storage and shipment.
INITIAL INSPECTION
Inspect the shipping container
(including cushioning material) for damage. If damaged,
keep
it until you have (1) checked the contents for completeness,
(2) read the three
following cautions and (3) checked the bridge mechanically and electrically. The
contents are listed in Table 2-1.
Table 2-1. Contents of lIP
85027
Directional Bridges
HP85027A
HP 85027B
HP 85027C
Instrument
Case
yes
yes
yes
Operating and
Service Manual yes
yes
yes
Adapter/
Connector saver
Type-N
m/Type-N m 3.5mm
3.5mm
m/3.Smm m
m/3.Smm f
Type-N m/Type-N m
Open/short
7mm open/short
3.5mm open/short
Type-N
Type-N
open
short
Wrench
no
yes
no
Read and observe these cautions: save yourself time and trouble.
CAUTION
Use caution when mating an SMA male connector to the
precision 3.5mm female connectors on the HP 85027B. Push
the connectors straight together. with the male contact concentric
with the female. DO NOT overtighten or rotate either center
conductor; turn only the outer nut of the male. An out of
spec
connector can permanently damage its mate. For
this reason, you
should measure
connectors with a connector gage (see Section 8,
“Mechanical Inspection”)
and use connector savers whenever
possible (see Figure 3-1).
CAUTION
Do not apply
more than
+23 dBm RF power or more than ±10
volts DC to the HP 85027.
More power or voltage will damage
the bridge.
2-1