HP 85027A/B/C Operation
OPERATOR’S CHECK
Figure 3-3 illustrates the setup for the operator’s check procedure. Follow this pro-
cedure to quickly check the entire measurement system.
Incorrect results may be caused
by any portion of the system, but if the HP 85027 is suspected use the performance tests
in Section 4 to determine whether the bridge is operating correctly. If the bridge fails
those tests, turn to
Section
8 to isolate the problem.
Equipment
POS Z
BLANK
P05
Z
BLANK
Figure 3-3. Typical Operator’s Check using HP 8757A
Bridge
Analyzer*
Sweep oscillator
RF plug-in
Calibrated open
Calibrated short
10 dB pad
HP 85027A
HP 8757A
HP 8350B
HP 83592A
85021-60001
85021-60001
8492A Opt 010
HP 85027B
HP 8757A
HP 8350B
HP 83595A
85037-60001
85037-60001
8493C Opt 010
HP 85027C
HP 8757A
HP 835013
HP 83592A
85032-60001
115 l2A
8491B Opt 010
*
ONLY REQUIRED
)F
HP 83508
HAS
SERIAL
PREFIX
2436 OR
BELOW.
*Note: If you perform this procedure with the HP 8756A or 8755C, (1) set the con-
figuration switch in step 2 to [HP8756/HP8755] and (2) do not perform step 9.
Proced umre
1.
Connect the equipment as shown in Figure 3-3 and turn it on.
2.
Set the HP
85027 switch (SI) to [HP8757].
3-4