HP 85027A/B/C Service
CAREFULLY ALIGN GAGE AND CON-
NECTOR SO THAT CIRCULAR BUSHING
AROUND GAGE PLUNGER WILL COME
TO REST ON OUTER CONDUCTOR
MATING PLANE.
ALLOWABLE RANGE
PROTRUSION
(—) 0.000 in
RECESSION (+) 0.003 in
Figure 8-7. Gaging Precision 3.5mm (m) Connector
OUTER CONDUCTOR
MATING PLANE
BUSHING
GRASP THE GAGE HERE
FOR MAXIMUM STABILITY.
OUTER CONDUCTOR
MATING PLANE
INSERT GAGE.
8-13