HP5501B
Service
8-27
8-81. If either the +15 Vdc or -15 Vdc 1.5 Amp fuse opens, the difference in potential that results
(across DS1 or DS2) will forward bias the associated LED indicator, turning that fuse indicator on.
8-82. System operating power is distributed to the laser head and other transducer units via the
connector board. Safety switch A1S1 opens when the laser head covers are removed. This
disconnects -15 Vdc from the high voltage power supply. As a result, this power supply
becomes inoperative. The +15 Vdc input is applied to regulator A3U7 on the A3
Reference/Control Board. This regulator pr5 Volts for use within the laser head. The +5
Volts is used to drive laser head's logic circuitry, READY LED (A1DS8), and as a reference for the
high voltage power supply switch.
8-83. A1 CONNECTOR BOARD TROUBLESHOOTING
HIGH VOLTAGE IS PRESENT ON A1J4 IF THE LASER IS ON AND
THE TUBE CONNECTOR IS NOT CONNECTED TO A1J4.
8-84. Use the schematic diagram Figure 8-23, an ohmmeter, and a voltmeter to troubleshoot A1.
Check to see if A1F1 is blown. Check for proper transmission of power supply voltages. Check
for proper diode action across A1CR1 and A1CR2. Check A1C1 through A1C5 for shorts. Check
switches A1S1 and A2S2. Check A1R3 and A1R4.
8-85. Rear Panel LED Operational Check
MAINTENANCE DESCRIBED HEREIN IS PERFORMED WITH
POWER SUPPLIED TO THE EQUIPMENT AND WITH THE
PROTECTIVE COVERS REMOVED. SUCH MAINTENANCE SHOULD
BE PERFORMED ONLY BY SERVICE- TRAINED PERSONNEL WHO
ARE AWARE OF THE HAZARDS INVOLVED (FOR EXAMPLE, FIRE
AND ELECTRIC SHOCK). WHERE MAINTENANCE CAN BE
PERFORMED WITHOUT POWER APPLIED, THE POWER TO THE
EQUIPMENT SHOULD BE REMOVED.
FROM 1.2 KILOVOLTS TO 12 KILOVOLTS DC IS PRESENT ON THE
ANODE OF THE LASER TUBE IN THE HP5501B. EXERCISE
EXTREME CAUTION WHEN WORKING INSIDE THE INSTRUMENT.
THE HIGH VOLTAGE COULD CAUSE SERIOUS PERSONAL INJURY
IF CONTACTED. THE INSTRUMENT SERVICE SHOULD BE
PERFORMED BY SERVICE-TRAINED PERSONNEL ONLY.
8-86. The following procedure can be used as an operational check of the rear panel LED’s. Perform
only if a problem is suspected.
a.
Remove
HP5501B
instrument covers as described in paragraph 8-55.
b.
Leave TEST-NORM switch (A1S2) set to NORM.
c.
Apply power to laser head as per Figure 4-1. (Use HP 6255A Dual DC Power Supply or
its equivalent.)
Содержание 5501B
Страница 2: ...ii ...
Страница 5: ...iii LASER HEAD SAFETY LABELS ...
Страница 12: ......
Страница 13: ...HP5501B Installation 1 1 Figure 1 1 HP5501B Laser Head ...
Страница 21: ...HP5501B Installation 2 3 10778B LENGTH 10 METRES 32 8 FT Figure 2 1 HP 10778B Power Cable ...
Страница 22: ...HP5501B Installation 2 4 10779B LENGTH 10 METRES 32 8 FT Figure 2 2 HP 10779B Reference Cable ...
Страница 46: ......
Страница 57: ...HP5501B Replaceable Parts 6 11 6 14 Exploded View Figure 6 1 HP5501B Laser Head Exploded View ...
Страница 60: ......
Страница 62: ...HP5501B Service 8 2 Figure 8 1 Schematic Diagram Notes ...
Страница 66: ...HP5501B Service 8 6 Figure 8 2 HP5501B Laser Head Block Diagram ...
Страница 91: ...HP5501B Service 8 31 Figure 8 18 High Voltage Power Supply Schematic Diagram ...
Страница 110: ...HP5501B Service 8 50 COMPONENT SIDE CIRCUIT SIDE SERIES 2632 REV A Figure 8 23 A1 Connector Board Component Locator ...
Страница 111: ...HP5501B Service 8 51 Figure 8 23 A1 Connector Board Component Schematic Diagram ...
Страница 118: ......