HP5501B
Operation
3-15
3-17. HP10746A Error Signal Processing
•
If any of the
HP10740A
backplane error bits are true, the four most significant bits of the
most significant word in a data transfer from the
HP10746A
to the host computer will be all
ones. These four bits are normally the decimal point bits.
•
Errors are not the only condition that can set these bits to all ones. When the difference
output of an addressed
HP10762A
is within tolerance, these bits will also be set.
•
After detecting all ones, the host computer must check the next four MSB's to differentiate
between an error condition and the within tolerance condition.
•
If the tolerance condition is true, these next four bits will also be all ones. Otherwise, each of
these bits corresponds to one of the backplane error bits.
3-18. HP 10745A Background Information
•
When a backplane error bit is true, the
HP10745A
asserts SRQ. The STATUS BYTE of the
HP10745A
can be read to determine the source of SRQ. I n the STATUS BYTE, bit 0
corresponds to the REFERENCE ERROR-BIT.
3-19. Program
Modifications
3-20. To monitor warmup time, the following program modifications can be added to the example
programs contained in Tables 4-12 and 4-13 of the
HP5501A
Laser Transducer Operating and
Service Manual, HP Part Number 05501-90028. If desired, some form of time-out algorithm can
be added to check if the
HP5501B
warms up within 10 minutes.
3-21. Table 3-1 shows the program steps used to modify Table 4-12 in the
HP5501A
System Manual.
The program uses the
HP10760A
x-axis card to monitor the laser warmup.
3-22. Table 3-2 shows the program steps used to modify Table 4-13 in the System Manual. The
program is for a two axis system with the
HP10746A
in 16-bit mode and using positive-true logic.
The program uses the x-axis card to monitor the laser warmup.
3-23. Tables 3-3, 3-4, and 3-5 are complete listings of the modified versions of Tables 4-12 and 4-13
found in the
HP5501A
System Manual.
Содержание 5501B
Страница 2: ...ii ...
Страница 5: ...iii LASER HEAD SAFETY LABELS ...
Страница 12: ......
Страница 13: ...HP5501B Installation 1 1 Figure 1 1 HP5501B Laser Head ...
Страница 21: ...HP5501B Installation 2 3 10778B LENGTH 10 METRES 32 8 FT Figure 2 1 HP 10778B Power Cable ...
Страница 22: ...HP5501B Installation 2 4 10779B LENGTH 10 METRES 32 8 FT Figure 2 2 HP 10779B Reference Cable ...
Страница 46: ......
Страница 57: ...HP5501B Replaceable Parts 6 11 6 14 Exploded View Figure 6 1 HP5501B Laser Head Exploded View ...
Страница 60: ......
Страница 62: ...HP5501B Service 8 2 Figure 8 1 Schematic Diagram Notes ...
Страница 66: ...HP5501B Service 8 6 Figure 8 2 HP5501B Laser Head Block Diagram ...
Страница 91: ...HP5501B Service 8 31 Figure 8 18 High Voltage Power Supply Schematic Diagram ...
Страница 110: ...HP5501B Service 8 50 COMPONENT SIDE CIRCUIT SIDE SERIES 2632 REV A Figure 8 23 A1 Connector Board Component Locator ...
Страница 111: ...HP5501B Service 8 51 Figure 8 23 A1 Connector Board Component Schematic Diagram ...
Страница 118: ......