HP5501B
Service
8-18
NOTE
Sampler Mount (MP1) is adjusted at the factory to ensure that the laser
beam is not truncated as it exits the laser head. Two screws (H5) secure
the Sampler Mount to the shield (MP10).
Figure 8-11. Factory-adjusted Sampler Mount (MP1)
8-60. To remove the A5 Laser Assembly (HP Part No. 05501-60102) from the
HP5501B
chassis,
perform the following procedure.
a.
Remove ±15 Volt power from
HP5501B
Laser Head.
b.
Remove front panel and side covers as described in paragraph 8-55.
c.
Disconnect laser assembly connector A5P1 from A1J4.
d.
Remove two Pozidriv screws (H7) that secures rear panel (MP4) to laser base (MP2).
Содержание 5501B
Страница 2: ...ii ...
Страница 5: ...iii LASER HEAD SAFETY LABELS ...
Страница 12: ......
Страница 13: ...HP5501B Installation 1 1 Figure 1 1 HP5501B Laser Head ...
Страница 21: ...HP5501B Installation 2 3 10778B LENGTH 10 METRES 32 8 FT Figure 2 1 HP 10778B Power Cable ...
Страница 22: ...HP5501B Installation 2 4 10779B LENGTH 10 METRES 32 8 FT Figure 2 2 HP 10779B Reference Cable ...
Страница 46: ......
Страница 57: ...HP5501B Replaceable Parts 6 11 6 14 Exploded View Figure 6 1 HP5501B Laser Head Exploded View ...
Страница 60: ......
Страница 62: ...HP5501B Service 8 2 Figure 8 1 Schematic Diagram Notes ...
Страница 66: ...HP5501B Service 8 6 Figure 8 2 HP5501B Laser Head Block Diagram ...
Страница 91: ...HP5501B Service 8 31 Figure 8 18 High Voltage Power Supply Schematic Diagram ...
Страница 110: ...HP5501B Service 8 50 COMPONENT SIDE CIRCUIT SIDE SERIES 2632 REV A Figure 8 23 A1 Connector Board Component Locator ...
Страница 111: ...HP5501B Service 8 51 Figure 8 23 A1 Connector Board Component Schematic Diagram ...
Страница 118: ......