HP5501B
Adjustments
5-5
h.
Apply power to the system power supplies.
NOTE
READY LED will flash rapidly for about 5 seconds. LASER LED
illuminates followed almost immediately by emission of laser light.
i.
Adjust A1R12 (Figure 5-1) for 3.5 milliamperes ± 0.1 milliampere. If the current cannot
be adjusted to this value, see Laser High Voltage Troubleshooting in
Section 8
. If the
laser power output (as measured in paragraph 4-9) is not greater than 120 microwatts
(at this current level), the laser assembly may require replacement.
j.
Record the laser tube current in the "ADJUSTMENT RECORD" located at the end of this
section.
k.
Turn off the system power supplies, disconnect the probe, set the TEST-NORM switch to
NORM, and reinstall the laser head covers.
l.
Return all A3 jumpers, moved in step 5-20(d), back to NRM position (left-most position).
Figure 5-3. 55018 High Voltage Areas and Anode Lead
Содержание 5501B
Страница 2: ...ii ...
Страница 5: ...iii LASER HEAD SAFETY LABELS ...
Страница 12: ......
Страница 13: ...HP5501B Installation 1 1 Figure 1 1 HP5501B Laser Head ...
Страница 21: ...HP5501B Installation 2 3 10778B LENGTH 10 METRES 32 8 FT Figure 2 1 HP 10778B Power Cable ...
Страница 22: ...HP5501B Installation 2 4 10779B LENGTH 10 METRES 32 8 FT Figure 2 2 HP 10779B Reference Cable ...
Страница 46: ......
Страница 57: ...HP5501B Replaceable Parts 6 11 6 14 Exploded View Figure 6 1 HP5501B Laser Head Exploded View ...
Страница 60: ......
Страница 62: ...HP5501B Service 8 2 Figure 8 1 Schematic Diagram Notes ...
Страница 66: ...HP5501B Service 8 6 Figure 8 2 HP5501B Laser Head Block Diagram ...
Страница 91: ...HP5501B Service 8 31 Figure 8 18 High Voltage Power Supply Schematic Diagram ...
Страница 110: ...HP5501B Service 8 50 COMPONENT SIDE CIRCUIT SIDE SERIES 2632 REV A Figure 8 23 A1 Connector Board Component Locator ...
Страница 111: ...HP5501B Service 8 51 Figure 8 23 A1 Connector Board Component Schematic Diagram ...
Страница 118: ......