Chapter 4 Command Reference
:TRIGger Subsystem
Programming Guide
4-43
4
:TRIGger Subsystem
COMMAND
:TRIGger[SEQuence | STARt]:HOLDoff . . . <numeric_value>
When INIT:CONT is ON, this command determines the rate at which
measurements are made by setting a delay between measurements. Its
setting corresponds to the front-panel rate setting as follows:
<NUMERIC_VALUE>
RATE
0.0 = FAST; 0.5 = MEDium; 1.0 = SLOW
COMMENTS
•
*RST: 0.0
•
The only settings that are accepted are the three listed above.
•
The measurement rate is also affected by the Resolution setting.
•
When INIT:CONT is OFF, the holdoff has no effect.
•
There is no setting for this command that corresponds to the front-
panel rate setting of HOLD. The HOLD mode is entered through *RST
or INIT:CONT OFF. The instrument then stops making
measurements until it is triggered or until it receives INIT:CONT ON.
RELATED
FRONT-PANEL
KEYS
Rate
Содержание 53150A Series
Страница 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Страница 2: ......
Страница 10: ...Contents x Programming Guide ...
Страница 11: ...1 Before You Start ...
Страница 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Страница 21: ...2 Command Summary A Quick Reference ...
Страница 41: ...3 Programming Your Counter for Remote Operation ...
Страница 100: ...Chapter 3 Programming Your Counter for Remote Operation Programming Examples 3 60 Programming Guide 3 ...
Страница 101: ...4 Command Reference A Dictionary ...
Страница 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Страница 163: ...5 Errors ...
Страница 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Страница 184: ......
Страница 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...