Chapter 4 Command Reference
:ABORt Command
4-4
Programming Guide
4
:ABORt Command
COMMAND
:ABORt
This command causes the Counter to abort, as quickly as possible, any
measurement in progress.
The :ABORt command is not complete until the current measurement is
stopped. The execution of an ABORt command sets false any Pending
Operation Flags that were set true by initiation of measuring.
COMMENTS
•
If :ABORt is issued while the measurement cycle is idle (:INIT:CONT
OFF and pending operation flag is false), the command is ignored.
•
If :ABORt is issued while a single measurement is in progress, the
measurement is aborted and the pending operation flag is set false.
•
If :ABORt is issued while repetitive measurements are being made
(:INIT:CONT ON), the measurement in progress is aborted, and the
pending operation flag is set false. Then, a new measurement is
automatically initiated, and the pending operation flag is set true.
•
If :ABORt is issued while a block of measurements, such as an
average, is in progress, the measurement block is aborted, and the
pending operation flag is set false.
•
When a measurement or block of measurements is aborted, the
Measuring bit in the Operation Status Register is set false.
•
Aborting a measurement in progress invalidates the result.
RELATED
FRONT-PANEL
KEYS
Reset/Local
Содержание 53150A Series
Страница 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Страница 2: ......
Страница 10: ...Contents x Programming Guide ...
Страница 11: ...1 Before You Start ...
Страница 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Страница 21: ...2 Command Summary A Quick Reference ...
Страница 41: ...3 Programming Your Counter for Remote Operation ...
Страница 100: ...Chapter 3 Programming Your Counter for Remote Operation Programming Examples 3 60 Programming Guide 3 ...
Страница 101: ...4 Command Reference A Dictionary ...
Страница 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Страница 163: ...5 Errors ...
Страница 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Страница 184: ......
Страница 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...