Chapter 4 Command Reference
Common Commands
4-52
Programming Guide
4
*PRE
(Parallel Poll Enable Register)
PRE?
(Parallel Poll Enable Register Query)
COMMAND
*PRE<NRf>
*PRE?
Sets or queries the value of the parallel poll enable register.
QUERY RESPONSE
<NRf>
RANGE
0-255
COMMENTS
•
The parallel poll enable register is eight bits wide and has the same
bit definitions as the status byte.
•
The status byte and parallel poll enable registers are anded together;
the result determines the value of *IST (TRUE or FALSE).
RELATED
FRONT-PANEL
KEYS
None
*RCL
(Recall Command)
COMMAND
*RCL <NRf> | <non-decimal numeric>
This command restores the state of the instrument from a copy stored in
local non-volatile memory. Before the recall occurs, the current state of
the instrument is automatically saved to register 0.
<NRF> RANGE
0 to 8
<NRF> RESOLUTION
1
RELATED
FRONT-PANEL
KEYS
Shift + Menu (RECALL > 0 through 8)
Содержание 53150A Series
Страница 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Страница 2: ......
Страница 10: ...Contents x Programming Guide ...
Страница 11: ...1 Before You Start ...
Страница 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Страница 21: ...2 Command Summary A Quick Reference ...
Страница 41: ...3 Programming Your Counter for Remote Operation ...
Страница 100: ...Chapter 3 Programming Your Counter for Remote Operation Programming Examples 3 60 Programming Guide 3 ...
Страница 101: ...4 Command Reference A Dictionary ...
Страница 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Страница 163: ...5 Errors ...
Страница 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Страница 184: ......
Страница 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...