9 Technical Data
82
9
Technical Data
9.1
Measurement Principle
9.1.1
Measurement principle using cross-modulation type NDIR
The molecular consisting of different atoms absorbs infrared ray of specific wavelength.
The cross-modulation type non-dispersive infrared analyzer detects the gap of infrared
absorption caused by alternatively introducing the sample gas and the reference gas by the
certain period to the measurement cell using the condenser microphone detector.
This method is free from zero drift. As for maintenance, the optical adjustments are not
required because it does not apply a rotation sector, which is quite effective for the stack gas
analyzers requiring continuous operation and high measurement accuracy.
Dividing the infrared ray from light source into two categories: reflected light and transmitted
light using a beam splitter allows the optical layout of as much as 4 components, realizing the
compact, downsizing module.
Fig. 64 Schematic diagram of measurement principle for NDIR
Light source
Beam splitter
SO
2
detector
NO detector
CO detector
CO
2
MAIN
MAIN
COMP
COMP
Subtraction
Subtraction
Subtraction
Subtraction
Measured value of SO
2
Measured value of NOx
Measured value of CO
2
Measured value of CO
Measurement
CO
2
cell
*1
COMP
MAIN
Sample gas
Reference gas
Solenoid valve
cell
sensor
*1