3 Basic Operation
11
3.3
Power Interruption
1:
Momentary power interruption for 0.02 seconds or less (Refer to 3.3.1.)
2:
Short power interruption occurring for 59 seconds or less which start and end within the
same minute of time (Refer to 3.3.2.)
3:
Short power interruption
occurring for 59 seconds or less which end at the next minute of
time (Refer to 3.3.3.)
4:
Long power interruption for more than 59 seconds (Refer to 3.3.3.)
3.3.1
Momentary power interruption
After the power restoration, ENDA-5000 returns to the state just before the power interruption.
If AIC or purge (optional) are in progress at the power interruption, AIC or purge continue to
start.
3.3.2
Power interruption 1
When AIC or purge (optional) are in progress at the power interruption, AIC or purge start from
the beginning of each sequence.
When the automatic calibration start time or purge start time has passed during the power
failure, the interval time is added to the start time when power is restored, and the time is
corrected into the future time nearest to the restoration time.
3.3.3
Power interruption 2
After the power restoration, the normal measurement condition is restored.
Note
When power failure occurs during “Battery” alarm, the internal clock, AIC start time, and the purge
start time is reset to 2090/1/1, 0:00 (initial value) after the power restoration.
12:00:00
12:01:00
12:02:00
12:04:00
12:05:00
12:03:00
12:06:00
Time
Energized status
Power interruption status
1
2
3
4