10 Optional Functions
102
Q
Mode
Sets the O
2
correction mode.
Pressing the button at the right of [MODE] in the O
2
CORRECTION SET screen displays
the O
2
CORRECTION MODE screen.
Fig. 77 O
2
CORRECTION MODE screen
Pressing the button of the item to be set, and press the [SET] key.
The setting is changed, and the screen returns to the O
2
CORRECTION SET screen.
Setting item
Description
EXTERNAL
Switches the O2 SET VALUE1 with O
2
SET VALUE 2 via external contact input
to perform the O
2
correction operation.
SET 1
Performs the O
2
correction operation using the value set by O
2
SET VALUE 1.
SET 2
Performs the O
2
correction operation using the value set by O
2
SET VALUE 2.